Repair GE 3SC-RS Crystal/Crystal Replacement/Damaged Crystals
Crystal failure reason
Most internal ultrasound probe uses a piezoelectric crystal. After long-term use, it can be causing the crystal of natural aging, and external force such as the use of non-standard or fall off, touch hit will make probe crystal damage, after crystal damaged, B ultrasonic detector detected in the image will appear signal attenuation, the image has dark, black bars, interference, defects such as blind area, serious the images will be "blind", which will cause doctors can not diagnose, affecting the quality of diagnosis.
Solution
This kind of failure is caused by ultrasound probe wafer short circuit. The short circuit of the chip can be judged by measuring the power supply voltage of the chip, if a chip or chip set of power supply voltage is lowerthan others, can judge the chip or chip set has a short circuit. Replace the corresponding chip or chip group, makes the probe to restore the previous feature, the interference disappears.
Other hot GE models we can repair:
GE | C1-5-D |
GE | C1-5-RS |
GE | 4C |
GE | 4C-RS |
GE | 4C-A |
GE | 4C-D |
GE | 4C-SC |
GE | E8C/E8C-RS |
GE | E8CS |
GE | RIC5-9-D |
GE | 3CRF-D |
GE | C1-6-D |
GE | RAB4-8-RS |
GE | RAB2-5-RS |
GE | 6S-RS |
GE | 12L-RS |
GE | 3SC-RS |
GE | L6-12-RS |
GE | 11L |
GE | 11L-D |
GE | 12L-RS |
GE | 8C |
GE | 3S |
GE | 3S-RS |
Common ultrasonic Probe Damage(Convex, Linear, Sector, Endocavity probes)
Common ultrasonic probe damage | Solutions |
Lens damage, wear, holes, swelling, delamination | Lens replacement |
Strain relief damage, separation | Strain replacement |
Nosepiece and probe separation and cracks |
Cosmetic repairs |
Cable cuts | Cable patches, possible cable replacement |
Connector housing electrical damage | Major and minor electrical repairs, pin module replacement |
Daily and Long-Term Storage of ultrasound transducer
1. Always store transducers in the transducer holders on the side of your system or on a securely mounted wall rack when you are not using them.
2. Ensure the transducer holders are clean before storing transducers
3. When storing transducers, use the cable-management clips, if available, to secure the transducer cable
4. Avoid storing transducers in areas of temperature extremes or in direct sunlight.
5. Store transducers separately from other instruments to avoid inadvertent transducer damage.
6. Before storing transducers, make sure they are thoroughly dry.